21 CFR 58.63: Maintenance and calibration of equipment
FDA's inspection data lists 1 Form 483 citation of 21 CFR 58.63 at 1 facility, from December 2014 to December 15, 2014.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 1
- At 1 inspection
- Facilities cited
- 1
- Distinct FEIs
- Most recent citation
- Dec 15, 2014
- Ascension Saint Thomas Hospital Midtown
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 58.63 from the Electronic Code of Federal Regulations.
§ 58.63 Maintenance and calibration of equipment.
(a) Equipment shall be adequately inspected, cleaned, and maintained. Equipment used for the generation, measurement, or assessment of data shall be adequately tested, calibrated and/or standardized.
(b) The written standard operating procedures required under § 58.81(b)(11) shall set forth in sufficient detail the methods, materials, and schedules to be used in the routine inspection, cleaning, maintenance, testing, calibration, and/or standardization of equipment, and shall specify, when appropriate, remedial action to be taken in the event of failure or malfunction of equipment. The written standard operating procedures shall designate the person responsible for the performance of each operation.
(c) Written records shall be maintained of all inspection, maintenance, testing, calibrating and/or standardizing operations. These records, containing the date of the operation, shall describe whether the maintenance operations were routine and followed the written standard operating procedures. Written records shall be kept of nonroutine repairs performed on equipment as a result of failure and malfunction. Such records shall document the nature of the defect, how and when the defect was discovered, and any remedial action taken in response to the defect.
21 CFR Part 58: Good Laboratory Practice for Nonclinical Laboratory Studies, Subpart DRead on eCFR
Citations by year
Form 483 citations of 21 CFR 58.63 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 1 in 2014
| Year | Form 483 citations |
|---|---|
| 2014 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 1 citation, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Equipment: maintenance SOPs1 citations
Citation text
- “The standard operating procedures for routine inspection, cleaning, maintenance, testing, calibration, and/or standardization of equipment are not adequate.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 58.63, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Often cited at the same inspection
Of the 1 inspection where FDA cited 21 CFR 58.63, the share where it also cited each of these sections.
- 21 CFR 606.100Standard operating procedures1 (100%)
Ask FDAgent about 21 CFR 58.63
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 58.63; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.