21 CFR 58.35: Quality assurance unit
FDA's inspection data lists 1 Form 483 citation of 21 CFR 58.35 at 1 facility, from March 2012 to March 7, 2012.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 1
- At 1 inspection
- Facilities cited
- 1
- Distinct FEIs
- Most recent citation
- Mar 7, 2012
- Bioclinica, Inc.
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 58.35 from the Electronic Code of Federal Regulations.
§ 58.35 Quality assurance unit.
(a) A testing facility shall have a quality assurance unit which shall be responsible for monitoring each study to assure management that the facilities, equipment, personnel, methods, practices, records, and controls are in conformance with the regulations in this part. For any given study, the quality assurance unit shall be entirely separate from and independent of the personnel engaged in the direction and conduct of that study.
(b) The quality assurance unit shall:
(1) Maintain a copy of a master schedule sheet of all nonclinical laboratory studies conducted at the testing facility indexed by test article and containing the test system, nature of study, date study was initiated, current status of each study, identity of the sponsor, and name of the study director.
(2) Maintain copies of all protocols pertaining to all nonclinical laboratory studies for which the unit is responsible.
(3) Inspect each nonclinical laboratory study at intervals adequate to assure the integrity of the study and maintain written and properly signed records of each periodic inspection showing the date of the inspection, the study inspected, the phase or segment of the study inspected, the person performing the inspection, findings and problems, action recommended and taken to resolve existing problems, and any scheduled date for reinspection. Any problems found during the course of an inspection which are likely to affect study integrity shall be brought to the attention of the study director and management immediately.
(4) Periodically submit to management and the study director written status reports on each study, noting any problems and the corrective actions taken.
(5) Determine that no deviations from approved protocols or standard operating procedures were made without proper authorization and documentation.
(6) Review the final study report to assure that such report accurately describes the methods and standard operating procedures, and that the reported results accurately reflect the raw data of the nonclinical laboratory study.
(7) Prepare and sign a statement to be included with the final study report which shall specify the dates inspections were made and findings reported to management and to the study director.
(c) The responsibilities and procedures applicable to the quality assurance unit, the records maintained by the quality assurance unit, and the method of indexing such records shall be in writing and shall be maintained. These items including inspection dates, the study inspected, the phase or segment of the study inspected, and the name of the individual performing the inspection shall be made available for inspection to authorized employees of the Food and Drug Administration.
(d) A designated representative of the Food and Drug Administration shall have access to the written procedures established for the inspection and may request testing facility management to certify that inspections are being implemented, performed, documented, and followed-up in accordance with this paragraph.
21 CFR Part 58: Good Laboratory Practice for Nonclinical Laboratory Studies, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 58.35 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 1 in 2012
| Year | Form 483 citations |
|---|---|
| 2012 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 1 citation, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- QAU: authorize deviations from protocols or SOPs1 citations
Citation text
- “The quality assurance unit failed to determine whether any deviations from approved protocols or standard operating procedures had been made with proper authorization and documentation.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 58.35, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Often cited at the same inspection
Of the 1 inspection where FDA cited 21 CFR 58.35, the share where it also cited each of these sections.
- 21 CFR 312.50General responsibilities of sponsors1 (100%)
Ask FDAgent about 21 CFR 58.35
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 58.35; FDAgent quotes that sentence exactly.