21 CFR 1271.10: Are my HCT/P's regulated solely under section 361 of the PHS Act and the regulations in this part, and if so what must I do?

FDA's inspection data lists no Form 483 citations of 21 CFR 1271.10. 4 warning letters state a violation of it.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
0
At 0 inspections
Facilities cited
0
Distinct FEIs
Most recent citation
—
None
Warning letters
4
State a violation of it

The regulation

The current text of 21 CFR 1271.10 from the Electronic Code of Federal Regulations.

§ 1271.10 Are my HCT/P's regulated solely under section 361 of the PHS Act and the regulations in this part, and if so what must I do?

(a) An HCT/P is regulated solely under section 361 of the PHS Act and the regulations in this part if it meets all of the following criteria:

(1) The HCT/P is minimally manipulated;

(2) The HCT/P is intended for homologous use only, as reflected by the labeling, advertising, or other indications of the manufacturer's objective intent;

(3) The manufacture of the HCT/P does not involve the combination of the cells or tissues with another article, except for water, crystalloids, or a sterilizing, preserving, or storage agent, provided that the addition of water, crystalloids, or the sterilizing, preserving, or storage agent does not raise new clinical safety concerns with respect to the HCT/P; and

(4) Either:

(i) The HCT/P does not have a systemic effect and is not dependent upon the metabolic activity of living cells for its primary function; or

(ii) The HCT/P has a systemic effect or is dependent upon the metabolic activity of living cells for its primary function, and:

(a) Is for autologous use;

(b) Is for allogeneic use in a first-degree or second-degree blood relative; or

(c) Is for reproductive use.

(b) If you are a domestic or foreign establishment that manufactures an HCT/P described in paragraph (a) of this section:

(1) You must register with FDA;

(2) You must submit to FDA a list of each HCT/P manufactured; and

(3) You must comply with the other requirements contained in this part.

21 CFR Part 1271: Human Cells, Tissues, and Cellular and Tissue-Based Products, Subpart ARead on eCFR

Warning letters stating a violation of 21 CFR 1271.10

4 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Your umbilical cord derived product fails to meet the criterion that the HCT/Ps be “intended for homologous use only.” Homologous use means that the “labeling, advertising, or other indications of the manufacturer’s objective intent” demonstrate that the HCT/P is intended to perform “the same basic function or functions in the recipient as in the donor” (21 CFR 1271.3(c) and 1271.10(a)(2)).”
“Your embryonic stem cell and placenta derived products fail to meet the criterion that the HCT/Ps be “intended for homologous use only.” Homologous use means that the “labeling, advertising, or other indications of the manufacturer’s objective intent” demonstrate that the HCT/P is intended to perform “the same basic function or functions in the recipient as in the donor” (21 CFR 1271.3(c) and 1271.10(a)(2)).”
“Your product fails to meet the criterion in 21 CFR 1271.10(a)(2) that the HCT/Ps be “intended for homologous use only.” Homologous use means that the “labeling, advertising, or other indications of the manufacturer’s objective intent” demonstrate that the HCT/P is intended to perform “the same basic function or functions in the recipient as in the donor” (21 CFR 1271.3(c) and 1271.10(a)(2)).”
“Your umbilical cord blood derived stem cell product fails to meet the criterion that the HCT/Ps be “intended for homologous use only.” Homologous use means that the “labeling, advertising, or other indications of the manufacturer’s objective intent” demonstrate that the HCT/P is intended to perform “the same basic function or functions in the recipient as in the donor” (21 CFR 1271.3(c) and 1271.10(a)(2)).”

Ask FDAgent about 21 CFR 1271.10

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 1271.10; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.